利用できるサービス | |
---|---|
修理 | 非対応 |
キャリブレーション | 対応 |
フリーサポート | 非対応 |
The CR23X Micrologger® was a portable, self-contained package that could store over 1,000,000 data points. It consisted of measurement and control electronics, communication ports, keyboard, 2-line alphanumeric display, power supply, and a battery-backed clock. The CR23X measured most sensor types directly, communicated via modems, reduced data, controlled external devices, and stored both data and programs in either non-volatile Flash memory or battery-backed SRAM.
続きを読むThe CR23X is a self-contained, rugged data logger that can measure most sensor types directly, communicate via modems, reduce data, control external devices, and store both data and programs in either non-volatile Flash memory or battery-backed SRAM.
The CR23X has an integral, 2-line, alphanumeric display and power supply. A battery-backed, real-time clock and 2M of nonvolatile data storage is included. The CR23Xs are available with a sealed rechargeable battery base, an alkaline base, or in a low-profile version without batteries.
-NOTE- | Note: Additional specifications are listed in the CR23X Specifications Sheet. |
Analog Inputs | 24 single-ended or 12 differential, individually configured |
Pulse Counters | 4 |
Switched Voltage Excitations | 4 |
Control/Digital I/O Ports | 8 |
Continuous Analog Outputs | 2 |
Serial I/O Port (CS I/O) | 1 |
RS-232 Port | 1 |
Scan Rate | 100 Hz |
Burst Mode | 1500 Hz |
Analog Voltage Resolution | To 0.33 uV |
A/D Bits | 15 |
Programming | Edlog |
Data Storage | Mixed Array, Table |
Telecommunications | Standard, Modbus, Alert |
Standard Operating Temperature Range | -25° to +50°C |
Extended Operating Temperature Range | -40° to +80°C |
Corrected Display Temperature | -20° to +40°C |
Survival Display Temperature | -40° to +85°C |
A typical field-based CR23X system consisted of:
Execution of this download installs the CR23X Operating System (Mixed-Array) on your computer.
Note: The Device Configuration Utility is used to upload the included operating system to the datalogger.
Execution of this download installs the CR23X Table Data Operating System on your computer.
Note: The Device Configuration Utility is used to upload the included operating system to the datalogger.
Execution of this download installs the CR23X PakBus Operating System on your computer.
Note: The Device Configuration Utility is used to upload the included operating system to the datalogger.
A software utility used to download operating systems and set up Campbell Scientific hardware. Also will update PakBus Graph and the Network Planner if they have been installed previously by another Campbell Scientific software package.
Supported Operating Systems:
Windows 11 or 10 (Both 32 and 64 bit)
CR23Xに関するよくある質問の数: 5
すべて展開すべて折りたたむ
E08 errors are occasionally created by voltage surges or transient voltages. Section 3.9 of the CR23X manual explains in greater detail what a watchdog is and possible causes. If these are happening frequently, it is likely an issue with the system that needs to be resolved. The first step in this resolution process would be to check the system power (inputs and outputs) and verify it is within the expected tolerances as outlined in the Specifications page (OV-23) of the CR23X Micrologger Operator's Manual.
バッテリ上部の刻印を探します。刻印は YYMMDDXX の日付形式になっている場合があります。
これはバッテリの寿命を示します。
Not the same, but similar. The SDM-SIO4 provides RS-232 voltage levels; the CR1000 control ports provide 0 to 5 V only. Both usually work with all sensors, and both devices are compatible with RS-232 and TTL logic. The CR1000 is easier to set up and program for serial input. The SDM-SIO1 is a preferred alternative to the SDM-SIO4.
When compared to the CR10X, the CR1000 can handle strings as a specific data type. It also has more integrated serial interfaces including the following:
The serial I/O capabilities of the CR1000/CR3000 are similar to, and faster than, the SDM-SIO4 capabilities on a CR10X or CR23X. SDM devices are addressable and are connected to a data logger on C1 through C3. Therefore, one benefit of using multiple SDM devices on a CR1000 datalogger is that only three control ports are used.
実用的な最大値は、データロガーの 2 つのCターミナルごとに 1 つのマルチプレクサを接続することです。Cターミナルをマルチプレクサ間で共有して、接続するマルチプレクサの数を増やすことができます。ただし、ターミナルを共有するには、より複雑な配線とプログラミングが必要です。2 つのCターミナルごとに 1 つ以上のマルチプレクサを接続したいユーザーは、Campbell Scientific の営業またはサポート エンジニアに問い合わせて支援を受けることをお勧めします。