Services Available | |
---|---|
Repair | No |
Calibration | No |
Free Support | No |
The AVW1 is available as a replacement part or as needed for continuity or compatibility, but Campbell Scientific recommends the AVW200-series interfaces for all new applications.
The AVW200 interfaces are much more accurate and feature a high degree of noise immunity for your vibrating wire measurements.
The AVW1 provides amplification and signal conditioning for the frequency signal of one vibrating wire or vibrating strip transducer. Compatible sensors include pressure, load, strain, or even vibrating wire temperature sensors. The recommended frequency range is 1000 to 9900 Hz (up to 5000 Hz in the CR5000).
続きを読むThe AVW1 is available as a replacement part or as needed for continuity or compatibility, but we recommend the AVW200-series interfaces for all new applications. The AVW200-series interfaces are much more accurate and feature a high degree of noise immunity for your vibrating wire measurements.
The AVW1 provides amplification and signal conditioning for a vibrating wire pressure, load, strain, or temperature sensor. The recommended frequency range is 1000 to 9900 Hz (up to 5000 Hz in the CR5000).
Frequency Sweep Range | 1000 to 9900 Hz (performance optimized for greater than 1000 Hz) |
Number of Vibrating Wire Sensors Measured | One vibrating wire sensor connects to the AVW1. Additional sensors can be measured by using an AM16/32-series multiplexer. |
注意: 以下は代表的な互換性情報を示しています。互換性のある製品や互換性のない製品をすべて網羅したリストではありません。
製品 | 互換性 | 注意 |
---|---|---|
21X (リタイア) | ||
CR10 (リタイア) | ||
CR1000 (リタイア) | ||
CR1000X | ||
CR10X (リタイア) | ||
CR200X (リタイア) | ||
CR206X (リタイア) | ||
CR211X (リタイア) | ||
CR216X (リタイア) | ||
CR23X (リタイア) | ||
CR295X (リタイア) | ||
CR3000 (リタイア) | ||
CR500 (リタイア) | ||
CR5000 (リタイア) | ||
CR510 (リタイア) | ||
CR800 (リタイア) | ||
CR850 (リタイア) | ||
CR9000 (リタイア) | ||
CR9000X (リタイア) |
Vibrating wire/strip transducers are measured by the VibratingWire Instruction in CRBasic, and by Instruction 28 (Vibrating Wire) in Edlog. The data logger produces a swept-frequency excitation, then measures a specified number of cycles of the sensor's resonant frequency.
AVW1に関するよくある質問の数: 2
The AVW1, AVW4, and AVW100 have a resolution of 0.1 Hz. Certifications of compliance for these items are not available.
The AVW1 is able to detect frequencies ranging from 1000 to 9900 Hz; the AVW100 is able to detect frequencies ranging from 250 to 9900 Hz.